Heroin (6-MAM) Test Evidence
Has Caused a False Positive
Nalorphine
Source Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The 2009 comparison lists weak reactivity for Nalorphine on the Microgenics CEDIA screen.
Keep in Mind
- This laboratory comparison used historical reagents.
- It gives no frequency after routine use.
- The result depends on the reagent, the cutoff and the concentration in the specimen.
Tests Studied
- Microgenics CEDIA Heroin metabolite (6-AM or 6-MAM) screen, Krasowski et al. supplement, H
- HEROIN METAB, row 17
Oxymorphone
Source Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The 2009 comparison lists weak reactivity for Oxymorphone on the Microgenics CEDIA screen.
Keep in Mind
- This laboratory comparison used historical reagents.
- It gives no frequency after routine use.
- The result depends on the reagent, the cutoff and the concentration in the specimen.
Tests Studied
- Microgenics CEDIA Heroin metabolite (6-AM or 6-MAM) screen, Krasowski et al. supplement, H
- HEROIN METAB, row 16
Depends on the Brand of Test
Codeine
Source Jenkins et al., J Anal Toxicol 2005
What It Found
- A 2005 CEDIA experiment found a positive heroin-marker screen at 60,000 ng/mL of codeine.
Source CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017
What It Found
- Later CEDIA instructions report a negative at 500,000 ng/mL.
Source Wang et al., J Anal Toxicol 2015
What It Found
- Recombinant-antibody HEIA remained negative at 1,000,000 ng/mL.
Test Studied
- Immunalysis recombinant-antibody 6-AM HEIA (10 ng/mL cutoff)
Source Cone et al., J Anal Toxicol 1991
Keep in Mind
- Controlled codeine administration did not produce 6-MAM by GC/MS.
Source Cone et al., J Anal Toxicol 1993
Keep in Mind
- A general opiates screen can detect codeine as expected, while a 6-MAM screen uses a distinct target.
2 sources Jenkins et al., J Anal Toxicol 2005 CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017
Test Studied
- Historical CEDIA 6-AM
2 sources Cone et al., J Anal Toxicol 1991 Cone et al., J Anal Toxicol 1993
Test Studied
- Specific urine GC/MS
Morphine
Source Jenkins et al., J Anal Toxicol 2005
What It Found
- A CEDIA study found that morphine at 10,000 ng/mL triggered the heroin-marker screen.
Source CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017
What It Found
- Later CEDIA instructions report a negative at 9,000 ng/mL.
Source Wang et al., J Anal Toxicol 2015
What It Found
- A recombinant-antibody test showed much weaker interference, with a response near cutoff around 300,000 ng/mL.
Test Studied
- Immunalysis recombinant-antibody 6-AM HEIA (10 ng/mL cutoff)
Source Yau et al., Forensic Sci Int 2012
What It Found
- The authors characterized the cross-reactivity of morphine and its 3- and 6-glucuronides as minimal on this reagent.
Keep in Mind
- The abstract does not state a morphine or glucuronide concentration that triggered a positive result.
Test Studied
- Emit II Plus 6-acetylmorphine on Viva-E
Source Moeller and Mueller, Forensic Sci Int 1995
What It Found
- In samples already containing 6-MAM by GC-MS, the authors attributed differences in RIA levels to morphine and morphine-6-glucuronide.
Keep in Mind
- The selected positive specimens demonstrate possible overestimation, without establishing a positive in target-free urine.
Test Studied
- Experimental 6-MAM radioimmunoassay
Source George et al., J Anal Toxicol 2002
What It Found
- All 21 unconfirmed positives in one series contained free morphine at 410–2,010 ng/mL.
Keep in Mind
- The authors could not distinguish morphine cross-reaction from another unidentified substance.
Test Studied
- Microgenics CEDIA 6-MAM versus GC-MS
Source Warren et al., Forensic Sci Int 2014
What It Found
- Morphine and both major morphine glucuronides showed minimal reactivity on the two ADVIA 6-AM applications.
Keep in Mind
- The abstract does not quantify the individual challenge thresholds.
- Do not transfer a historical CEDIA result to this reagent.
Test Studied
- Emit II Plus 6-AM on Siemens ADVIA 1800 and 2400
Source Heideloff et al., Clin Chem 2014
Keep in Mind
- A lab also reported a separate, rare sample-preparation artifact with high-morphine urine during enzymatic hydrolysis.
- A disputed low result warrants method review by the laboratory.
4 sources Smith et al., Forensic Sci Int 2014 Jenkins et al., J Anal Toxicol 2005 CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
Test Studied
- Historical CEDIA 6-AM
Poppy seeds
Source Smith et al., Forensic Sci Int 2014
What It Found
- The older CEDIA heroin-marker test produced false positives in some urine collected after controlled poppy seed ingestion.
- The Lin-Zhi 6-AM assay produced none.
- Investigators measured genuine morphine and codeine from the seeds.
Test Studied
- Thermo CEDIA Heroin Metabolite versus Lin-Zhi 6-AM, 10 ng/mL cutoff
Both sources Smith et al., Forensic Sci Int 2014 Cone et al., J Anal Toxicol 1991
Keep in Mind
- 6-acetylmorphine (6-AM, 6-MAM or 6-monoacetylmorphine) is a different molecule.
- A food-related morphine result does not account for 6-MAM confirmed on a suitably specific method.
Hydrocodone
Source Jenkins et al., J Anal Toxicol 2005
What It Found
- A 2005 study found a positive CEDIA heroin-marker screen with a hydrocodone standard at 60,000 ng/mL.
Source CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017
What It Found
- Later CEDIA instructions report a negative at 300,000 ng/mL.
Source Wang et al., J Anal Toxicol 2015
What It Found
- Recombinant-antibody HEIA remained negative at 1,000,000 ng/mL.
Test Studied
- Immunalysis recombinant-antibody 6-AM HEIA (10 ng/mL cutoff)
All 3 sources Jenkins et al., J Anal Toxicol 2005 CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017 Wang et al., J Anal Toxicol 2015
Keep in Mind
- These are concentrations in laboratory specimens.
- They do not establish how often an ordinary prescription causes a positive urine screen or explain a confirmed 6-MAM result.
2 sources Jenkins et al., J Anal Toxicol 2005 CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017
Test Studied
- Historical CEDIA 6-AM
Hydromorphone
Source Jenkins et al., J Anal Toxicol 2005
What It Found
- A 2005 study found a positive CEDIA heroin-marker screen with a hydromorphone standard at 10,000 ng/mL.
Source CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017
What It Found
- Later CEDIA instructions report a negative at the same concentration.
Source Wang et al., J Anal Toxicol 2015
What It Found
- The recombinant-antibody HEIA study reported a near-cutoff response only at a much higher level.
Test Studied
- Immunalysis recombinant-antibody 6-AM HEIA (10 ng/mL cutoff)
All 4 sources Jenkins et al., J Anal Toxicol 2005 CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017 Wang et al., J Anal Toxicol 2015 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
Keep in Mind
- These are concentrations in laboratory specimens.
- They do not establish how often an ordinary prescription causes a positive urine screen or explain a confirmed 6-MAM result.
Oxycodone
Source Jenkins et al., J Anal Toxicol 2005
What It Found
- A 2005 study found a positive CEDIA heroin-marker screen with a oxycodone standard at 61,000 ng/mL.
Source CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017
What It Found
- Later CEDIA instructions report a negative at 400,000 ng/mL.
Source Wang et al., J Anal Toxicol 2015
What It Found
- Recombinant-antibody HEIA remained negative at 1,000,000 ng/mL.
Test Studied
- Immunalysis recombinant-antibody 6-AM HEIA (10 ng/mL cutoff)
All 3 sources Jenkins et al., J Anal Toxicol 2005 CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017 Wang et al., J Anal Toxicol 2015
Keep in Mind
- These are concentrations in laboratory specimens.
- They do not establish how often an ordinary prescription causes a positive urine screen or explain a confirmed 6-MAM result.
2 sources Jenkins et al., J Anal Toxicol 2005 CEDIA Heroin Metabolite (6-AM) Assay, manufacturer instructions 10006723-10-EN, August…, 2017
Test Studied
- Historical CEDIA 6-AM
Pentazocine
Source Jenkins et al., J Anal Toxicol 2005
What It Found
- A 2005 study found a positive CEDIA heroin-marker screen with a pentazocine standard at 35,000 ng/mL.
Test Studied
- Historical CEDIA 6-AM
Source Wang et al., J Anal Toxicol 2015
What It Found
- Wang and colleagues included pentazocine among compounds with no interference on the recombinant-antibody assay.
Test Studied
- Immunalysis recombinant-antibody 6-AM HEIA (10 ng/mL cutoff)
Both sources Jenkins et al., J Anal Toxicol 2005 Wang et al., J Anal Toxicol 2015
Keep in Mind
- These are concentrations in laboratory specimens.
- They do not establish how often an ordinary prescription causes a positive urine screen or explain a confirmed 6-MAM result.
The Test Is Made to Find It
Heroin
Source Cone et al., J Anal Toxicol 1991
What It Found
- Heroin breaks down to 6-acetylmorphine, also called 6-MAM, 6-AM or 6-monoacetylmorphine.
- This metabolite was detected after heroin administration but not after morphine or codeine in controlled testing.
Keep in Mind
- A specific confirmation carries a different meaning from a preliminary heroin-marker screen.
- The metabolite clears rapidly, so its absence does not exclude an earlier exposure.
Test Studied
- Targeted urine 6-acetylmorphine GC/MS after controlled administration