Hydrocodone Test Evidence
Depends on the Brand of Test
Codeine
Both sources Nascimento et al., J Anal Toxicol 2016 Oyler et al., J Anal Toxicol 2000
What It Found
- There are two possible reasons for a positive.
Keep in Mind
- Specific testing can distinguish cross-reactivity from the presence of hydrocodone.
Source Nascimento et al., J Anal Toxicol 2016
What It Found
- Codeine itself reacted with the Lin-Zhi test in spiked urine.
Test Studied
- Lin-Zhi hydrocodone/hydromorphone HEIA, 100 and 300 ng/mL cutoffs
Source Oyler et al., J Anal Toxicol 2000
What It Found
- In a separate controlled dosing study, people taking codeine produced small amounts of actual hydrocodone in urine.
Keep in Mind
- When hydrocodone is present, the laboratory should also consider its amount relative to codeine and the medication history.
Morphine
Source Nascimento et al., J Anal Toxicol 2016
What It Found
- Added morphine reacted on the Lin-Zhi hydrocodone/hydromorphone screen.
Test Studied
- Lin-Zhi hydrocodone/hydromorphone HEIA, 100 and 300 ng/mL cutoffs
Source Cone et al., J Anal Toxicol 2008
What It Found
- Separately, people taking high doses of morphine have excreted small amounts of hydromorphone, a target of this combined screen.
Keep in Mind
- A confirmed small amount of hydromorphone can reflect morphine metabolism.
- This does not mean that morphine becomes hydrocodone.
Both sources Nascimento et al., J Anal Toxicol 2016 Cone et al., J Anal Toxicol 2008
Keep in Mind
- Look at the specific analytes measured.
Dihydrocodeine
Source Nascimento et al., J Anal Toxicol 2016
What It Found
- Dihydrocodeine at 5,000 ng/mL triggered the 100 ng/mL cutoff.
- It did not trigger the 300 ng/mL cutoff even at 20,000 ng/mL.
Keep in Mind
- A cutoff-specific cross-reaction demonstrated in the laboratory does not establish the frequency at prescribed doses.
Test Studied
- Lin-Zhi hydrocodone/hydromorphone HEIA
A Related Drug the Test Picks Up
Oxycodone
Source Nascimento et al., J Anal Toxicol 2016
What It Found
- Added oxycodone produced a positive at 5,000 ng/mL on the 100 ng/mL cutoff and 6,000 ng/mL on the 300 ng/mL cutoff.
Test Studied
- Lin-Zhi hydrocodone/hydromorphone HEIA, 100 and 300 ng/mL cutoffs
Source Dixon and Dasgupta, Ther Drug Monit 2016
What It Found
- Of the other opioids challenged at 5,000 ng/mL, only oxycodone and oxymorphone gave even minimal cross-reactivity.
Keep in Mind
- A measurable signal was reported.
- The abstract does not show that this signal reached the decision cutoff.
Test Studied
- DRI hydrocodone/hydromorphone assay, 100 ng/mL decision cutoff
Both sources Nascimento et al., J Anal Toxicol 2016 Dixon and Dasgupta, Ther Drug Monit 2016
Keep in Mind
- This is a hydrocodone/hydromorphone-specific screen.
- Drug identification by follow-up testing must account for real metabolites as well as antibody cross-reactions.
Oxymorphone
Source Nascimento et al., J Anal Toxicol 2016
What It Found
- Added oxymorphone produced a positive at 2,000 ng/mL on the 100 ng/mL cutoff and 5,000 ng/mL on the 300 ng/mL cutoff.
Test Studied
- Lin-Zhi hydrocodone/hydromorphone HEIA, 100 and 300 ng/mL cutoffs
Source Dixon and Dasgupta, Ther Drug Monit 2016
What It Found
- Of the other opioids challenged at 5,000 ng/mL, only oxycodone and oxymorphone gave even minimal cross-reactivity.
Keep in Mind
- A measurable signal was reported.
- The abstract does not show that this signal reached the decision cutoff.
Test Studied
- DRI hydrocodone/hydromorphone assay, 100 ng/mL decision cutoff
Both sources Nascimento et al., J Anal Toxicol 2016 Dixon and Dasgupta, Ther Drug Monit 2016
Keep in Mind
- This is a hydrocodone/hydromorphone-specific screen.
- Drug identification by follow-up testing must account for real metabolites as well as antibody cross-reactions.
The Test Is Made to Find It
Hydrocodone
Source Nascimento et al., J Anal Toxicol 2016
What It Found
- An assay for hydrocodone and hydromorphone can appropriately detect prescribed hydrocodone and its metabolite.
Tests Studied
- Hydrocodone/hydromorphone HEIA
- UPLC-MS/MS
Source Stolbach et al., J Med Toxicol 2022
Keep in Mind
- A generic opiate assay has different sensitivity.
- Determine which test was performed.
Hydromorphone
Source Nascimento et al., J Anal Toxicol 2016
What It Found
- This particular reagent is designed to recognize hydrocodone and hydromorphone.
- A person prescribed hydromorphone can therefore screen positive.
Keep in Mind
- A positive on the combined assay cannot distinguish prescribed hydromorphone from hydromorphone produced by metabolism of hydrocodone.
Test Studied
- Lin-Zhi hydrocodone/hydromorphone HEIA