Back to the drug test guide

Hydrocodone Test Evidence

Depends on the Brand of Test

Codeine

Both sources Nascimento et al., J Anal Toxicol 2016 Oyler et al., J Anal Toxicol 2000

What It Found

  • There are two possible reasons for a positive.

Keep in Mind

  • Specific testing can distinguish cross-reactivity from the presence of hydrocodone.

Source Nascimento et al., J Anal Toxicol 2016

What It Found

  • Codeine itself reacted with the Lin-Zhi test in spiked urine.

Test Studied

  • Lin-Zhi hydrocodone/hydromorphone HEIA, 100 and 300 ng/mL cutoffs

Source Oyler et al., J Anal Toxicol 2000

What It Found

  • In a separate controlled dosing study, people taking codeine produced small amounts of actual hydrocodone in urine.

Keep in Mind

  • When hydrocodone is present, the laboratory should also consider its amount relative to codeine and the medication history.

Morphine

Source Nascimento et al., J Anal Toxicol 2016

What It Found

  • Added morphine reacted on the Lin-Zhi hydrocodone/hydromorphone screen.

Test Studied

  • Lin-Zhi hydrocodone/hydromorphone HEIA, 100 and 300 ng/mL cutoffs

Source Cone et al., J Anal Toxicol 2008

What It Found

  • Separately, people taking high doses of morphine have excreted small amounts of hydromorphone, a target of this combined screen.

Keep in Mind

  • A confirmed small amount of hydromorphone can reflect morphine metabolism.
  • This does not mean that morphine becomes hydrocodone.

Both sources Nascimento et al., J Anal Toxicol 2016 Cone et al., J Anal Toxicol 2008

Keep in Mind

  • Look at the specific analytes measured.

Dihydrocodeine

Source Nascimento et al., J Anal Toxicol 2016

What It Found

  • Dihydrocodeine at 5,000 ng/mL triggered the 100 ng/mL cutoff.
  • It did not trigger the 300 ng/mL cutoff even at 20,000 ng/mL.

Keep in Mind

  • A cutoff-specific cross-reaction demonstrated in the laboratory does not establish the frequency at prescribed doses.

Test Studied

  • Lin-Zhi hydrocodone/hydromorphone HEIA

A Related Drug the Test Picks Up

Oxycodone

Source Nascimento et al., J Anal Toxicol 2016

What It Found

  • Added oxycodone produced a positive at 5,000 ng/mL on the 100 ng/mL cutoff and 6,000 ng/mL on the 300 ng/mL cutoff.

Test Studied

  • Lin-Zhi hydrocodone/hydromorphone HEIA, 100 and 300 ng/mL cutoffs

Source Dixon and Dasgupta, Ther Drug Monit 2016

What It Found

  • Of the other opioids challenged at 5,000 ng/mL, only oxycodone and oxymorphone gave even minimal cross-reactivity.

Keep in Mind

  • A measurable signal was reported.
  • The abstract does not show that this signal reached the decision cutoff.

Test Studied

  • DRI hydrocodone/hydromorphone assay, 100 ng/mL decision cutoff

Both sources Nascimento et al., J Anal Toxicol 2016 Dixon and Dasgupta, Ther Drug Monit 2016

Keep in Mind

  • This is a hydrocodone/hydromorphone-specific screen.
  • Drug identification by follow-up testing must account for real metabolites as well as antibody cross-reactions.

Oxymorphone

Source Nascimento et al., J Anal Toxicol 2016

What It Found

  • Added oxymorphone produced a positive at 2,000 ng/mL on the 100 ng/mL cutoff and 5,000 ng/mL on the 300 ng/mL cutoff.

Test Studied

  • Lin-Zhi hydrocodone/hydromorphone HEIA, 100 and 300 ng/mL cutoffs

Source Dixon and Dasgupta, Ther Drug Monit 2016

What It Found

  • Of the other opioids challenged at 5,000 ng/mL, only oxycodone and oxymorphone gave even minimal cross-reactivity.

Keep in Mind

  • A measurable signal was reported.
  • The abstract does not show that this signal reached the decision cutoff.

Test Studied

  • DRI hydrocodone/hydromorphone assay, 100 ng/mL decision cutoff

Both sources Nascimento et al., J Anal Toxicol 2016 Dixon and Dasgupta, Ther Drug Monit 2016

Keep in Mind

  • This is a hydrocodone/hydromorphone-specific screen.
  • Drug identification by follow-up testing must account for real metabolites as well as antibody cross-reactions.

The Test Is Made to Find It

Hydrocodone

Source Nascimento et al., J Anal Toxicol 2016

What It Found

  • An assay for hydrocodone and hydromorphone can appropriately detect prescribed hydrocodone and its metabolite.

Tests Studied

  • Hydrocodone/hydromorphone HEIA
  • UPLC-MS/MS

Source Stolbach et al., J Med Toxicol 2022

Keep in Mind

  • A generic opiate assay has different sensitivity.
  • Determine which test was performed.

Hydromorphone

Source Nascimento et al., J Anal Toxicol 2016

What It Found

  • This particular reagent is designed to recognize hydrocodone and hydromorphone.
  • A person prescribed hydromorphone can therefore screen positive.

Keep in Mind

  • A positive on the combined assay cannot distinguish prescribed hydromorphone from hydromorphone produced by metabolism of hydrocodone.

Test Studied

  • Lin-Zhi hydrocodone/hydromorphone HEIA