PCP Test Evidence
Has Caused a False Positive
Mesoridazine
Source Long et al., Clin Chem 1996
What It Found
- A patient’s urine screened positive for PCP by EMIT II and was negative by GC-MS.
- Both thioridazine and its metabolite mesoridazine were present.
Keep in Mind
- The case cannot identify which compound produced the signal.
Test Studied
- EMIT II PCP in a patient’s urine
2 sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- Separate specificity data found cross-reactivity to mesoridazine on three assays.
Keep in Mind
- Laboratory thresholds varied from 24 to 400 micrograms/mL depending on the reagent.
Test Studied
- Historical Abbott, Syva and Biosite specificity data
Methylenedioxypyrovalerone
Source Macher and Penders, Drug Test Anal 2013
What It Found
- A published report attributed false-positive PCP immunoassay results to MDPV, a synthetic cathinone.
Keep in Mind
- The indexed letter has no abstract.
- This is a specific report, and the method details and frequency are unverified here.
Test Studied
- Urine PCP immunoassay in a published report
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Methylenedioxypyrovalerone challenge, Microgenics DRI Phencyclidine gave a positive result at 3.5 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Thioridazine
Source Long et al., Clin Chem 1996
What It Found
- A patient’s urine screened positive for PCP by EMIT II and was negative by GC-MS.
- Both thioridazine and its metabolite mesoridazine were present.
Keep in Mind
- The urine case cannot distinguish the parent medicine from its metabolite.
Test Studied
- EMIT II PCP in a patient’s urine
2 sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- In separate manufacturer data, thioridazine reacted on Abbott AxSYM at a very high concentration.
Keep in Mind
- The AxSYM concentration does not establish the likelihood at a prescribed dose.
Test Studied
- Historical Abbott AxSYM specificity data
Tramadol
2 sources Ly et al., Ann Emerg Med 2012 Hull et al., Am J Forensic Med Pathol 2006
What It Found
- Several overdose cases produced PCP-positive urine.
- Laboratory experiments with tramadol and its metabolites supported cross-reactivity.
Keep in Mind
- Concentrations in overdose reports can be much higher than with typical dosing.
Test Studied
- Siemens Syva EMIT II and EMIT II Plus PCP screens
Source Rengarajan and Mullins, Clin Toxicol (Phila) 2013
What It Found
- A separate emergency department study found tramadol was associated with unconfirmed screens.
Venlafaxine
Source Bond et al., J Toxicol Clin Toxicol 2003
What It Found
- A venlafaxine overdose produced a positive Abbott AxSYM result with no PCP by GC-MS.
- Adding venlafaxine and its active metabolite to urine reproduced the result.
Keep in Mind
- The Abbott spiking experiment used very high concentrations.
Test Studied
- Abbott AxSYM PCP
3 sources Sena et al., Clin Chem 2002 Kim et al., Case Rep Psychiatry 2023 Landy and Kripalani, BJPsych Bull 2015
What It Found
- Separate reports describe possible interference during prescribed use.
Test Studied
- Other rapid urine PCP immunoassays
Source Landy and Kripalani, BJPsych Bull 2015
What It Found
- Two patients taking prescribed venlafaxine repeatedly screened positive for PCP in a rehabilitation center.
Keep in Mind
- The two rehabilitation cases had other prescriptions, and the authors did not report mass-spectrometric confirmation for either.
Desvenlafaxine
Source Farley et al., BMJ Case Rep 2017
What It Found
- A case report linked a presumptive PCP result with prescribed desvenlafaxine.
Keep in Mind
- The prescription case described the link as likely.
- Independent confirmation was unavailable in that case.
Test Studied
- Medtox PCP urine screen
Both sources Farley et al., BMJ Case Rep 2017 Sena et al., Clin Chem 2002
What It Found
- Desvenlafaxine is also the active metabolite implicated in studies of venlafaxine.
Lamotrigine
All 4 sources Geraci et al., Int J Emerg Med 2010 Murphy et al., J Emerg Nurs 2019 McCain et al., Prim Care Companion CNS Disord 2018 Cole, Am J Emerg Med 2022
What It Found
- Several reports describe positive rapid PCP screens in people taking lamotrigine.
Keep in Mind
- Confirmation varied between reports.
Source Geraci et al., Int J Emerg Med 2010
What It Found
- In an early two-patient report, broader testing was negative in one patient and the other was assessed from history.
Test Studied
- Bio-Rad TOX/See
3 sources Murphy et al., J Emerg Nurs 2019 McCain et al., Prim Care Companion CNS Disord 2018 Cole, Am J Emerg Med 2022
What It Found
- Further reports describe the finding on other rapid systems.
Source Cole, Am J Emerg Med 2022
Keep in Mind
- The 2022 report used blood for confirmation of a urine result and involved other medications.
4-MePPP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-MePPP challenge, Microgenics DRI Phencyclidine gave a positive result at 18.5 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-MPBP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-MPBP challenge, Microgenics DRI Phencyclidine gave a positive result at 9 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-MeO-DALT
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-MeO-DALT challenge, Microgenics DRI Phencyclidine gave a positive result at 78 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Alpha-PBP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Alpha-PBP challenge, Microgenics DRI Phencyclidine gave a positive result at 4.6 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Alpha-PPP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Alpha-PPP challenge, Microgenics DRI Phencyclidine gave a positive result at 7 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Alpha-PVP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Alpha-PVP challenge, Microgenics DRI Phencyclidine gave a positive result at 1.45 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Alpha-PVT
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Alpha-PVT challenge, Microgenics DRI Phencyclidine gave a positive result at 0.6 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
D2PM
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the D2PM challenge, Microgenics DRI Phencyclidine gave a positive result at 65 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
MDPBP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the MDPBP challenge, Microgenics DRI Phencyclidine gave a positive result at 24 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
MDPPP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the MDPPP challenge, Microgenics DRI Phencyclidine gave a positive result at 60 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Meperidine
2 sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The assay comparison reported meperidine cross-reactivity at high urine concentrations.
Keep in Mind
- Cross-reactivity depends on the assay and concentration.
Test Studied
- Historical Abbott ARCHITECT and Syva EMIT PCP assays
Source Krasowski et al., BMC Emerg Med 2009
What It Found
- One unconfirmed Syva EMIT positive contained high levels of both meperidine and dextromethorphan.
Keep in Mind
- The patient sample does not identify which drug caused the signal.
Source Parsons et al., Clin Chem 1993
What It Found
- The original Advisor study obtained a positive PCP result from Meperidine at 100 mg/L of drug-free urine, with no reaction at 10 mg/L.
Keep in Mind
- These were fortified samples and high laboratory concentrations.
- Ordinary-dose clinical detection was not tested.
Test Studied
- Abbott Advisor PCP urine immunoassay
Methoxetamine
Source Gomila et al., J Anal Toxicol 2019
What It Found
- The ketamine analog methoxetamine showed weak cross-reactivity in a laboratory comparison of PCP screens.
Keep in Mind
- Reactivity was only 0.04% to 0.25%, and the investigators emphasized limited detectability.
Test Studied
- Five commercial PCP immunoassays
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Methoxetamine challenge, Microgenics DRI Phencyclidine gave a positive result at 90 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Metronidazole
Source Fernández et al., Clin Toxicol (Phila) 2018
What It Found
- A report identified metronidazole interference on a PCP screen.
Both sources Fernández et al., Clin Toxicol (Phila) 2018 Saitman et al., J Anal Toxicol 2026
What It Found
- Added metronidazole produced a positive response in urine on the tested device.
Keep in Mind
- The laboratory concentration was high.
- This finding does not establish a probability during usual treatment.
Test Studied
- One-step multi-analyte urine drug screen
MOPPP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the MOPPP challenge, Microgenics DRI Phencyclidine gave a positive result at 26 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
MPHP (study designation)
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the MPHP challenge, Microgenics DRI Phencyclidine gave a positive result at 1.7 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Naphyrone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Naphyrone challenge, Microgenics DRI Phencyclidine gave a positive result at 11 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Pyrovalerone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Pyrovalerone challenge, Microgenics DRI Phencyclidine gave a positive result at 2 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Trihexyphenidyl
Source Parsons et al., Clin Chem 1993
What It Found
- The original Advisor study obtained a positive PCP result from Trihexyphenidyl at 100 mg/L of drug-free urine, with no reaction at 10 mg/L.
Keep in Mind
- These were fortified samples and high laboratory concentrations.
- Ordinary-dose clinical detection was not tested.
Test Studied
- Abbott Advisor PCP urine immunoassay
Depends on the Brand of Test
Dextromethorphan
Source Krasowski et al., BMC Emerg Med 2009
What It Found
- Eight unconfirmed Syva EMIT positives in one series had high urine dextromethorphan concentrations.
Keep in Mind
- Overdose-level exposure was prominent.
Source Rengarajan and Mullins, Clin Toxicol (Phila) 2013
What It Found
- A separate emergency department study found dextromethorphan was more often recorded when a positive PCP screen failed GC-MS confirmation.
Source Schier, J Emerg Med 2000
What It Found
- A separate original letter describes a dextromethorphan-associated presumptive PCP urine result.
Keep in Mind
- The citation does not supply the assay, dose or confirmatory findings.
Test Studied
- Published letter concerning a urine PCP drug screen
Source Lin et al., J Anal Toxicol 2013
What It Found
- The authors observed no interference from Dextromethorphan at 1,000 ng/mL when it was added to drug-free urine for the cup comparison.
Keep in Mind
- This is a single challenge level.
- It cannot rule out interference at a higher concentration or with a different device.
Test Studied
- NexScreen and DrugCheck Waive RT cups, as applicable to their respective panels
2 sources Krasowski et al., BMC Emerg Med 2009 Rengarajan and Mullins, Clin Toxicol (Phila) 2013
Keep in Mind
- Routine cough dosing and other manufacturers’ screens may give different results.
Test Studied
- Syva EMIT and Siemens Syva EMIT II PCP screens
Also cited: Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
Diphenhydramine
Source Krasowski et al., BMC Emerg Med 2009
What It Found
- One Triage-positive specimen contained a high urine diphenhydramine concentration without GC-MS evidence of PCP.
Keep in Mind
- The stronger clinical lead involved very high exposure.
Source Rengarajan and Mullins, Clin Toxicol (Phila) 2013
What It Found
- In another study diphenhydramine was more common in unconfirmed screens, but the difference was statistically inconclusive.
Source Lin et al., J Anal Toxicol 2013
What It Found
- The authors observed no interference from Diphenhydramine at 1,000 ng/mL when it was added to drug-free urine for the cup comparison.
Keep in Mind
- This is a single challenge level.
- It cannot rule out interference at a higher concentration or with a different device.
Test Studied
- NexScreen and DrugCheck Waive RT cups, as applicable to their respective panels
2 sources Krasowski et al., BMC Emerg Med 2009 Rengarajan and Mullins, Clin Toxicol (Phila) 2013
Keep in Mind
- The frequency at usual doses is unknown.
Test Studied
- Biosite Triage and Siemens Syva EMIT II PCP screens
Ketamine
Source Shannon, Pediatr Emerg Care 1998
What It Found
- An early clinical letter reported a PCP-positive screen after ketamine administration.
Test Studied
- Older clinical urine toxicology screen
Source Gomila et al., J Anal Toxicol 2019
What It Found
- In a later laboratory evaluation, ketamine did not react on any of five PCP immunoassays.
- Some related designer drugs did react.
Test Studied
- Five commercial PCP immunoassays
Both sources Shannon, Pediatr Emerg Care 1998 Gomila et al., J Anal Toxicol 2019
Keep in Mind
- Ketamine itself has poor support as a general explanation.
- Check the exact test and avoid applying data on designer analogues to prescribed ketamine.
2-DPMP
Source Beck et al., Drug Test Anal 2014
What It Found
- At a Desoxypipradrol spike of 4,500 ng/mL, the mean assay response was 26 ng/mL equivalent (0.6% cross-reactivity), above the 25 ng/mL cutoff.
Keep in Mind
- This is a mean laboratory signal at the stated concentration.
- The separate clinical urine findings can include metabolites or coexposures.
Test Studied
- CEDIA Phencyclidine urine assay, cutoff 25 ng/mL
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2-DPMP challenge, Microgenics DRI Phencyclidine gave a positive result at 21 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
A Related Drug the Test Picks Up
3-Methoxyphencyclidine
Source Gomila et al., J Anal Toxicol 2019
What It Found
- This PCP analog reacted on all five assays in laboratory testing and in urine from two people with confirmed analog overdose.
Keep in Mind
- A PCP-class screen can respond to closely related substances.
- It cannot determine which molecule is present.
Test Studied
- Five commercial PCP immunoassays
4-Methoxyphencyclidine
Source Gomila et al., J Anal Toxicol 2019
What It Found
- This PCP analog reacted on all five assays in an experimental comparison.
Keep in Mind
- A PCP-class screen can respond to closely related substances.
- It cannot determine which molecule is present.
Test Studied
- Five commercial PCP immunoassays
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-Methoxyphencyclidine challenge, Microgenics DRI Phencyclidine gave a positive result at 1.5 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Benocyclidine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Benocyclidine challenge, Microgenics DRI Phencyclidine gave a positive result at 0.4 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Suspected, but Not Proven
Alprazolam
Source Rengarajan and Mullins, Clin Toxicol (Phila) 2013
What It Found
- In an emergency department chart study, this medicine was more commonly recorded among PCP screens that failed GC-MS confirmation.
Keep in Mind
- The authors noted an association with polysubstance use.
- The analysis did not isolate this medicine as an analytical cross-reactant.
Test Studied
- Siemens Syva EMIT II PCP
Carvedilol
Source Rengarajan and Mullins, Clin Toxicol (Phila) 2013
What It Found
- In an emergency department chart study, this medicine was more commonly recorded among PCP screens that failed GC-MS confirmation.
Keep in Mind
- The authors noted an association with polysubstance use.
- The analysis did not isolate this medicine as an analytical cross-reactant.
Test Studied
- Siemens Syva EMIT II PCP
Clonazepam
Source Rengarajan and Mullins, Clin Toxicol (Phila) 2013
What It Found
- In an emergency department chart study, this medicine was more commonly recorded among PCP screens that failed GC-MS confirmation.
Keep in Mind
- The authors noted an association with polysubstance use.
- The analysis did not isolate this medicine as an analytical cross-reactant.
Test Studied
- Siemens Syva EMIT II PCP
Ibuprofen
Source Martínez Sánchez et al., An Pediatr (Barc) 2011
What It Found
- A child who had taken therapeutic ibuprofen had a positive MPD Screen 12 result.
- GC-MS detected ibuprofen and excluded PCP.
Keep in Mind
- The child also took lamotrigine, carbamazepine and two cough preparations.
Test Studied
- MPD Screen 12 Test panel (Linear Chemicals)
Source Marchei et al., Ther Drug Monit 2007
What It Found
- A separate Instant-View report involved an overdose with ibuprofen, metamizole and dextromethorphan.
Test Studied
- Instant-View multi-test
Both sources Marchei et al., Ther Drug Monit 2007 Martínez Sánchez et al., An Pediatr (Barc) 2011
Keep in Mind
- Neither case isolates ibuprofen as the cause of the false positive.
Metamizole
Source Marchei et al., Ther Drug Monit 2007
What It Found
- A letter investigated an overdose involving ibuprofen, metamizole and dextromethorphan in connection with a PCP-positive screen.
Keep in Mind
- The accessible report metadata do not establish an independent effect of this ingredient.
- Dextromethorphan has separate evidence and attribution should remain open.
Test Studied
- Instant-View multi-test in an overdose report
Tested: No False Positive Found
Doxylamine
Source Schaldenbrand et al., Ther Drug Monit 1981
What It Found
- In a doxylamine intoxication, the PCP immunoassay was negative.
- A separate gas chromatographic method falsely identified PCP.
Keep in Mind
- This commonly cited case cannot establish a false positive on an immunoassay.
- The chromatographic method is distinct from contemporary tandem MS confirmation.
Test Studied
- EMIT-dau urine immunoassay and historical gas chromatography without mass spectrometry
2,5-Dimethoxyamphetamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2,5-Dimethoxyamphetamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
2-Aminoindane
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2-Aminoindane challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
25B-NBOMe
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 25B-NBOMe challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
25C-NBOMe
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 25C-NBOMe challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
25I-NBOMe
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 25I-NBOMe challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
2C-B
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2C-B challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
2C-C
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2C-C challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
2C-E
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2C-E challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
2C-I
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2C-I challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
2C-T
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2C-T challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
2C-T-2
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2C-T-2 challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
2C-T-4
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2C-T-4 challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
2C-T-7
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 2C-T-7 challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
3,4,5-Trimethoxyamphetamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 3,4,5-Trimethoxyamphetamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
3,4-Dimethylmethcathinone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 3,4-Dimethylmethcathinone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
3-Trifluoromethylphenylpiperazine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 3-TFMPP challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-CAP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-CAP challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-EEC
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-EEC challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-Fluoroamphetamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-Fluoroamphetamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-Fluoromethamphetamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-Fluoromethamphetamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-HO-DET
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-HO-DET challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-MEC
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-MEC challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-Methylamphetamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-Methylamphetamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-Methylthioamphetamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-Methylthioamphetamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
4-TFMPP
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 4-TFMPP challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-APB
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-APB challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-APDB
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-APDB challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-APDI
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-APDI challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-IAI
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-IAI challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-IT
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-IT challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-MeO-AMT
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-MeO-AMT challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-MeO-DiPT
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-MeO-DiPT challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-MeO-DMT
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-MeO-DMT challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
5-MeO-MiPT
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 5-MeO-MiPT challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
6-APB
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the 6-APB challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Acetaminophen
Source Lin et al., J Anal Toxicol 2013
What It Found
- The authors observed no interference from Acetaminophen at 1,000 ng/mL when it was added to drug-free urine for the cup comparison.
Keep in Mind
- This is a single challenge level.
- It cannot rule out interference at a higher concentration or with a different device.
Test Studied
- NexScreen and DrugCheck Waive RT cups, as applicable to their respective panels
Alpha-methyltryptamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Alpha-methyltryptamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Benzedrone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Benzedrone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Benzylpiperazine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Benzylpiperazine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Bromo-DragonFLY
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Bromo-DragonFLY challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Brompheniramine
Source Thermo Scientific. DRI Phencyclidine Assay instructions for use, 0141, 2025
What It Found
- Adding 50,000 ng/mL of brompheniramine did not produce a positive PCP result on this assay.
Keep in Mind
- Brompheniramine can affect some amphetamines screens.
- That association does not establish a false-positive PCP result.
Test Studied
- Thermo Scientific DRI Phencyclidine urine assay, 25 ng/mL cutoff
Buphedrone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Buphedrone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Butylone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Butylone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Cathinone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Cathinone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Dibenzylpiperazine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Dibenzylpiperazine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Diethylpropion
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Diethylpropion challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Dimethylamylamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Dimethylamylamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Dimethylone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Dimethylone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Dimethyltryptamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Dimethyltryptamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Dipropyltryptamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Dipropyltryptamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
DiPT
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the DiPT challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
DOB
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the DOB challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
DOC
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the DOC challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
DOET
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the DOET challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
DOI
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the DOI challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
DOM
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the DOM challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Escaline
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Escaline challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Ethcathinone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Ethcathinone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Ethylone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Ethylone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Eutylone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Eutylone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Flephedrone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Flephedrone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Imipramine
Source Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- Five PCP screening reagents in the published comparison showed no reaction to imipramine.
Keep in Mind
- The table does not give challenge concentrations for those five reagents.
- It cannot predict the result on another PCP assay.
Test Studied
- Abbott Architect, Beckman, Biosite Triage, Microgenics DRI and Roche Cobas c PCP screens, supplement tab P, row 34
MDAI
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the MDAI challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
MDEA
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the MDEA challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Mephedrone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Mephedrone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Meprobamate
Source Lin et al., J Anal Toxicol 2013
What It Found
- The authors observed no interference from Meprobamate at 1,000 ng/mL when it was added to drug-free urine for the cup comparison.
Keep in Mind
- This is a single challenge level.
- It cannot rule out interference at a higher concentration or with a different device.
Test Studied
- NexScreen and DrugCheck Waive RT cups, as applicable to their respective panels
Mescaline
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Mescaline challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Meta-chlorophenylpiperazine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Meta-chlorophenylpiperazine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Methcathinone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Methcathinone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Methedrone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Methedrone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Methiopropamine
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Methiopropamine challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Methylone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Methylone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
MT-45
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the MT-45 challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
N,N-Dimethylcathinone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the N,N-Dimethylcathinone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Pentedrone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Pentedrone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Pentylone
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the Pentylone challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
PMA
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the PMA challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
PMMA
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the PMMA challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine
Prednisone
Source FDA 510(k) K253567. Healgen Accurate Urine Drug Screen Dip Card and Home Urine Drug Test…, 2025
What It Found
- Prednisone at 100 µg/mL did not cause a positive result in drug-free urine or change the result of urine just below or above this channel’s cutoff.
Keep in Mind
- The manufacturer challenged the assay with parent prednisone.
- This does not test all prednisone metabolites or predict the behavior of other devices.
Tests Studied
- Healgen Accurate multi-drug urine dip card
- The applicable target and cutoff are 25 ng/mL phencyclidine
WIN 35,428
Source Regester et al., J Anal Toxicol 2015
What It Found
- In the WIN 35,428 challenge, Microgenics DRI Phencyclidine stayed negative at 100 µg/mL.
Keep in Mind
- These were pure reference compounds added to drug-free human urine.
- The concentrations do not predict a positive result after use in a patient.
Test Studied
- Urine screening on Roche/Hitachi Modular P, Microgenics DRI Phencyclidine