Propoxyphene Test Evidence
Has Caused a False Positive
Ranitidine
Source Ayala-Lopez et al., J Anal Toxicol 2022
What It Found
- Adding ranitidine to urine at 800 micrograms/mL made the dedicated propoxyphene strip positive.
Keep in Mind
- This incidental high-concentration finding is specific to one point-of-care cup.
Test Studied
- Alere Integrated E-Z Split Key Cup II, propoxyphene strip
6-Acetylcodeine
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list 6-Acetylcodeine as weakly reactive at 60 micrograms/mL.
Keep in Mind
- 6-acetylcodeine is a separate substance.
- Taking prescribed codeine is not evidence of this exposure.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
Bepridil
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list Bepridil as weakly reactive at 100 micrograms/mL.
Keep in Mind
- These figures came from historical reagent information compiled in 2009.
- They do not establish risk after usual doses.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
Cetirizine
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list Cetirizine as weakly reactive at 50 micrograms/mL.
Keep in Mind
- These figures came from historical reagent information compiled in 2009.
- They do not establish risk after usual doses.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
Fexofenadine
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list Fexofenadine as weakly reactive at 20 micrograms/mL.
Keep in Mind
- These figures came from historical reagent information compiled in 2009.
- They do not establish risk after usual doses.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
Hydroxyzine
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list Hydroxyzine as weakly reactive at 30 micrograms/mL.
Keep in Mind
- These figures came from historical reagent information compiled in 2009.
- They do not establish risk after usual doses.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
Indomethacin
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list Indomethacin as weakly reactive at 70 micrograms/mL.
Keep in Mind
- These figures came from historical reagent information compiled in 2009.
- They do not establish risk after usual doses.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
Nortriptyline
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list 10-hydroxynortriptyline, a nortriptyline metabolite, as weakly reactive at 100 micrograms/mL.
Keep in Mind
- These figures came from historical reagent information compiled in 2009.
- They do not establish risk after usual doses.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
Oxybutynin
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list Oxybutynin as weakly reactive at 60 micrograms/mL.
Keep in Mind
- These figures came from historical reagent information compiled in 2009.
- They do not establish risk after usual doses.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
Depends on the Brand of Test
Diphenhydramine
Source Schneider and Wennig, J Anal Toxicol 1999
What It Found
- A laboratory letter reported EMIT II interference.
Test Studied
- Historical EMIT II
2 sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- A 2009 compilation listed weak Biosite Triage reactivity at 70 micrograms/mL but no reaction on several other evaluated reagents, including the Syva reagent in its comparison.
Tests Studied
- Biosite Triage PPX
- Comparisons with other propoxyphene reagents
Source Poklis et al., J Anal Toxicol 2004
What It Found
- Diphenhydramine cross-reacted in urine during specificity testing of Triage PPY.
Keep in Mind
- This experiment does not establish a positive at a usual dose or on other propoxyphene assays.
Test Studied
- BIOSITE TRIAGE Plus Propoxyphene (TPP), 300 ng/mL target cutoff
All 4 sources Schneider and Wennig, J Anal Toxicol 1999 Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009 Poklis et al., J Anal Toxicol 2004
Keep in Mind
- Separate primary-report and package-insert findings differ.
- Device, generation, concentration and metabolites may matter.
- A personal risk at ordinary doses is unknown.
Clemastine
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list Clemastine as weakly reactive at 40 micrograms/mL.
- The same compilation lists no reaction for Roche Cobas c.
Keep in Mind
- These figures came from historical reagent information compiled in 2009.
- They do not establish risk after usual doses.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
Cyclobenzaprine
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The 2009 manufacturer-data review lists weak reactivity at 99 micrograms/mL on Abbott and 80 on Syva.
- Biosite Triage did not react under the reported conditions.
Keep in Mind
- Package-insert data describe those reagent versions and challenge concentrations.
- Patient event rates were not determined.
Test Studied
- Historical Abbott Architect and Syva EMIT propoxyphene reagents, 300 ng/mL cutoff
Imipramine
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The 2009 manufacturer-data review lists weak reactivity at 40 micrograms/mL on Abbott and 30 on Syva.
- Beckman, Biosite Triage, Microgenics DRI and Roche Cobas c did not react under the reported conditions.
Keep in Mind
- Package-insert data describe those reagent versions and challenge concentrations.
- Patient event rates were not determined.
Test Studied
- Historical Abbott Architect and Syva EMIT propoxyphene reagents, 300 ng/mL cutoff
Methaqualone
Both sources Krasowski et al., BMC Emerg Med 2009 Krasowski et al. Additional file 1. Historical assay cross-reactivity tables and reagent…, 2009
What It Found
- The compiled manufacturer data list Methaqualone as weakly reactive at 100 micrograms/mL.
- Abbott Architect, Beckman, Microgenics DRI, Roche and Syva were listed as unreactive.
Keep in Mind
- These figures came from historical reagent information compiled in 2009.
- They do not establish risk after usual doses.
Test Studied
- Biosite Triage PPX, historical 300 ng/mL cutoff
The Test Is Made to Find It
Propoxyphene
Source Kintz and Mangin, J Anal Toxicol 1993
What It Found
- This test detected propoxyphene, and measured its genuine major metabolite norpropoxyphene with 29.3–92.6% cross-reactivity depending on concentration.
Keep in Mind
- Validation included urine and other biological specimens.
- The metabolite percentages characterize the assay, not detection rates in patients.
Test Studied
- Abbott TDx propoxyphene FPIA and GC-MS